MRI Safety and Policies & Procedures
MRI Safety and Policies & Procedures
Magnet Safety at ALL TIMES
Outline
Understanding Magnets
o Your role in MR Safety
o Metallic Screening
o Screening Patients / Colleagues
o Other Safety Considerations
What to do in Emergencies
MRI Department Policies and Procedures
Preview MRI Safety Videotape
Magnetism / Magnets
* All substances possess some form of magnetism.
* The degree of magnetism exhibited depends on the atoms that make-up the substance.
* Magnetic susceptibility is the ability of a substance to become magnetized.
* Ferromagnetic substances, such as iron have a large magnetic susceptibility, it is easily magnetized permanently and becomes a magnet itself.
* All magnets have a North and a South pole.
* All magnets have a “fringe” magnetic field which exists in the vicinity surrounding the magnet.
Magnetic Fringe Fields
* The fringe magnetic field is the magnetic field which exists in the vicinity surrounding the magnet.
* This field may extend many meters from the magnet itself.
* These imaginary lines of force demonstrate the pattern of the magnetic field.
* Safety and operational concerns make it necessary to contain the fringe field to a small area.
* Magnetic fields are measured in units of Gauss or Tesla.
MRI Safety at ALL TIMES
A STATIC MAGNETIC FIELD IS ALWAYS PRESENT 24hrs/day, 365 days/yr. EVEN WHEN NOT IN USE.
ANY PERSON USING THE MAGNET MUST BE CERTIFIED AFTER ATTENDING THE MRI SAFETY TRAINING CLASS.
ONE MUST BE TRAINED ON THE SCANNER INTERFACE BEFORE SCANNING.
What is your role in MR Safety?
The greatest risk of injury and damage to the system results from:
o Misuse or abuse of the MR equipment
o Failure to comply with recommended safety procedures
o Lack of proper inspection and maintenance of the MR equipment
Who should know MR Safety?
All in-house personnel that have reason to enter the MR suite area should be trained in MR safety procedures:
o MR technologists, students, researchers, transporters and other medical personnel
o Maintenance and janitorial personnel
All personnel must be thoroughly briefed about the potential risks involved and reminded not to bring any ferromagnetic items into the magnetic field.
Examples of items in-house personnel may have that can become projectiles if brought into the magnetic field
Tool Kits
Clipboards, Metal Pens
Tools
Gurneys, Wheelchairs
Vacuum Cleaners
Oxygen Cylinders
Buffers
Stethoscopes
Buckets
Scalpels, Syringes, Needles
Dustpans
Scissors, Hemostats
Maintenance & Janitorial personnel
MR technologists, students, researchers, transporters & other medical personnel
Who should know MR Safety?
Public safety forces that may respond to the MR suite for an emergency must also know the potential hazards of the MR equipment.
o Law enforcement personnel
o Fire department personnel
A person from the MR site should discuss the possible hazards with these people and provide them with handouts that will reinforce the information.
Examples of items public safety forces may have that can become projectiles if brought into the magnetic field
Breathing apparatus
Fire Extinguishers
Flashlights
Pike poles
Clipboards
Nozzles
Handcuffs
Hose couplings
Knives
Axes
Guns
Fire Department personnel
Law enforcement personnel
MRI Safety - Projectiles
* Projectile effects of metal objects seriously compromise safety. The potential harm cannot be over emphasized.
* Many types of clinical equipment are ferromagnetic and should never be brought into the scan room.
* Items may be tested for magnetic susceptibility with a hand-held magnet located at each MR station.
Metal Objects Becoming Projectiles
Fatal Accidents CAN Happen!
Patient Emergencies
Should a condition exist where the patient is having a medical emergency, all efforts must be made to quickly and safely remove the patient from the scan room.
Once the patient is removed from the MR scan room, close the door to prevent re-entry.
Under no circumstances should a “code team” be allowed to enter the scan room without proper screening!
Controlled Access Area
Although not detectable by the human senses, a magnetic field can be dangerous to equipment and to people.
Since a magnet is always “at field,” safety procedures must be followed to prevent accidents.
For the safety of patients and personnel, controlled access areas are established.
Controlled Access Area
* These areas are established for the safety of patients and personnel.
* The area is labeled with the use of warning signs and markings to prevent the entry of ferromagnetic objects into the controlled access area and to limit the access of individuals with medical implants near high magnetic fields.
* Public access begins at the 5 gauss line (0.5mT).
Equipment / Personal Items
The magnetic field can seriously damage or impair the operation of equipment or personal items such as:
o Oscilloscopes (slow moving electron beams)
o Camera
o Watches
o Credit / Bank cards
o Hearing Aids
o Hair Accessories, Belt Buckles, Shoes
Screening Procedures
* At least one MR operator must screen the patient for possible contraindications that could affect the MR scan. See Patient History and Safety Screening form.
* Check implanted devices in the Reference Manual for Magnetic Resonance Safety by Frank Shellock, Ph.D. or by using MRIsafety.com
Example of MRI Metal Screening Sheet
Screening Procedures
Static magnetic fields can alter the operation of electrically and mechanically operated implants and must remain outside the 5 gauss line.
Pregnant medical personnel should take precautions and remain outside of the magnet room during scanning.
Questions about implants not found in reference material should be discussed with a licensed, MRI technologist or a radiologist before allowing the patient to be scanned.
Absolute Contraindications
* Cardiac Pacemakers (except in rare, controlled environments)
* Cochlear (inner ear) implants
* Swan-Ganz catheters with thermodilution tips
* Ferromagnetic or unidentifiable aneurysm clips of the brain
* Implanted neuro stimulators
* Metal or unidentifiable foreign bodies in the eyes
* Shrapnel near a vital organ
Bioeffects
There is no conclusive evidence for irreversible or harmful bioeffects in humans below 3.0T.
Reversible abnormalities may include but are not limited to:
o Localized tissue and core body temperature heating
o Cutaneous sensations (tingling)
o Peripheral nerve stimulation (involuntary muscle contractions)
o Burn hazards
* Burn Hazards are caused by damaged hardware or by electrical currents produced in conductive loops of material.
* Localized heating is caused by RF irradiation energy absorption to a volume of tissue.
* Dissipation of the absorbed RF energy is described in terms of Specific Absorption Rate (SAR), measured in watts/kg.
* SAR is calculated by the patient’s weight and the expected increase in body temperature for each imaging pulse sequence.
* Patients with poor thermo-regulatory systems must be carefully monitored.
Acoustic Noise
The MR scanner can produce very high acoustic noise levels.
Some patients may experience discomfort from the associated noise of the scanner.
Prior to scanning, it is strongly recommended that earplugs be provided to the patient to reduce the noise level by at least 25dB.
Operating Safely
When operating the MR equipment, be attentive to the following abnormal conditions:
* Louder-than-normal motor noises
* Sparks
* Components overheating
* Smoke or odors coming from the electronic equipment or from within the scan room.
Do not operate equipment with protective panels opened or removed, there is risk of electric shock and can cause image artifacts.
Magnetic Field / Scan Room Emergencies
If an emergency situation arises, you may need to quickly bring down the patient systems and remove power from the MR system.
The nature of the emergency will dictate which procedure you follow. Each procedure has a distinct and specific purpose.
Each magnet is equipped with two emergency buttons:
* Emergency Stop / Shut Off
o Turns off all incoming electrical power to the magnet Power Distribution Unit (PDU)
* Quench or Emergency Run Down
o Causes immediate collapse of the superconductive magnetic field within minutes
FAMILIARIZE YOURSELF WITH THESE BUTTONS. KNOW THE DIFFERENCE!
Emergency Stop / Shut Off Button
Shutting power to the PDU may be required for life threatening situations such as:
* Fire in the computer room
* Fire, sparks, loud noises emanating from the scan room
* Flooding or sprinkling system goes off
* Catastrophic equipment failure
***Keep in mind that when this button is pushed, it does not initiate a quench, the magnet remains “at field.” Exercise caution, make sure that all ferromagnetic materials remain outside of the scan room***
Quench / Emergency Run Down Button
The following situation is THE ONLY TIME that may require quenching of the magnet:
* Large magnetic object pins or impales a person against the magnet and no other method can prevent further injury or free the person.
Do not attempt to pull large magnetic objects (oxygen tanks) from a magnet field. The object may change its magnetic polarity and re-align itself on the magnet and become a projectile, causing a serious or fatal injury.
Do not touch a quenched magnet. Under certain conditions, an electrical potential of >1,000 volts could exist on the surface of the magnet.
Quenching
Definition: a loss of superconductivity of the magnet coil due to a local temperature increase in the magnet as it becomes resistive, resulting in rapid evaporation of liquid helium in the cryostat and quickly reducing the magnetic field strength.
* A quench may happen spontaneously or can be manually instigated in case of an emergency.
* Quenching may cause severe and irreparable damage to the superconducting coils (magnet).
* A magnet quench will result in several days’ downtime, so do not press the button except in a true emergency.
* Do not attempt to test this button!
Emergency Buttons @ MR1 Univ. of Utah Hospital
* QUENCH BUTTON
o Button is located on the east wall (with window).
* E-STOP BUTTON
o Button is located behind the door as you enter the scan room (on the right).
QUENCH
OXYGEN
SENSOR
QUENCH
University of Utah Hospital and Clinics
MRI Department Policies and Procedures
This manual is available at all sites having a MRI scanner. Detail of all departmental situations can be reviewed. The following safety considerations are further highlighted:
Cryogen Safety Oxygen Monitors
Metallic Screening Pregnancy / Nursing
Magnet Quench Medical Emergencies
Magnetic Field / Scan Room Emergencies
Summary
* MRI scanners are powerful magnets with the ability to attract ferromagnetic objects.
* Any personnel around the MRI suite must be adequately screened for metallic implants and personal items before entering the scan room.
* Patients in the scanner must be carefully monitored for reversible bioeffects caused by the magnet’s hardware.
* Become familiarized with E-Stop vs. Quench buttons at each scanner.
* Review Policies and Procedures Manual
Congratulations!
You have completed the University of Utah Hospital and Clinics MRI Safety Training course!
Please review a safety video that demonstrates the powerful forces of MRI magnets.
Following written certification, you will be authorized to aid or assist an MRI technologist with patient examination procedures.
MRI Safety and Policies & Procedures.ppt
0 comments:
Post a Comment