25 April 2009

Magnetoencephalography



Magnetoencephalography (MEG )

How It Works

* Currents in neurons create very tiny magnetic fields
* MEG uses SQUIDs to detect these magnetic fields.
* Magnetic signals from the brain are only a few fT in strength.
* Needs a magnetically shielded room


History

* First measured by David Cohen in 1968.
* He used a copper induction coil
* Presently, MEG technology uses SQUIDS
* Today, MEG machines can contain as many as 300 SQUID sensors


Diagram of MEG setup

SQUIDS

* Superconducting Quantum Interference Devices
* Superconducting material is niobium or lead alloy with gold/indium
* Cooled to low temperature with either liquid He (4K) or N (77K)
* Manufactured at NIST in Boulder!
* Noise levels about 3 fT*Hz^-1/2

Josephson Junctions

* Two superconductors separated by a thin insulating barrier
* A small current will tunnel across the barrier
* The constant current Ic depends on temperature and magnetic field
* SQUIDS measure fractions of the phase difference in terms of the flux quantum h/2e

Detecting Brain Activity

* 50,000 neurons need to fire to generate a readable signal
* Neurons near the outside of the brain generate the strongest signals

Magnetic Shielding
Forward Problem
The Inverse Problem
Synthetic Aperture Magnetometry
Magnetic Source Imaging
Dipole Model Source Localization
Lead-field-based imaging approach
Independent Component Analysis
Uses of MEG

* MEGs are used in research to measure the time course of brain activity
* MEGs can detect epilepsy, as well as detect areas of the brain that are most important to avoid during surgery

Advantages/Disadvantages

* High 1 ms time resolution
* Completely non-invasive
* Does not depend on head geometry like EEG
* Magnetic fields decay faster over distance than electric fields
* MEG is best used to complement other imaging techniques


References
Magnetoencephalography (MEG).ppt

0 comments:

All links posted here are collected from various websites. No video or powerpoint files are uploaded on this blog. If you are the original author and do not wish to display your content on this blog please Email me anandkumarreddy at gmail dot com I will remove it. The contents of this blog are meant for educational purpose and not for commercial use. If you use any content give due credit to the original author.